Kowa fm600 Inflatable Bouncy Toy User Manual


 
35
Scanning laser source
Photoreceiver element
Measurement range
Measurement accuracy
Measurement area
Measurement time
Measurement distance
Printer
Monitor
Energy saving function
Interface
Power supply
Dimension/Weight
Semiconductor laser, Wave length: 635nm, Intensity: 35±15µW
A photomultiplier tube
1 - 500 Photon count/ms
±5%(under the calibration environment)
0.3mm (H) X 0.5mm (W)
0.5 seconds
81mm (Patient’s eye - objective lens)
58mm - wide thermal printer
5.6" TFT color LCD
Lengthwise: 37mm, Crosswise 89mm, Vertical: 25mm
Provided
Input: AC100V - 230V, 50/60Hz, power consumption: 80VA (Normal), 100VA (Max)
274(W) X 457(D) X 458(H)mm / 18 Kg
12. Specifications
12-1 Specifications
12-2 Completed safety standards and Classification
1. IEC 60601-1:1988+A1:1991+A2:1995
Classification of the medical device according to on the
protection type against electrical shock
Classification of the patient attachment according to on the
protection level against electrical shock
Classification based on IEC60529: the level of protection
that electrical appliances provide against the intrusion of
water.
Classification of safe use of the device based on the level
of protection against the environment which contains air
and flammable anesthesia gases, or the environment
which contains oxygen or nitrous-oxide gas and flammable
anesthesia gases
Classification of the device based on the operation mode
Class I
Type B
IPXO
This is not a device capable of being used in such
environment.
A device capable of being operated continuously
Complied EC Directive
Meidcal Device Directive 93/42/EEC
WEEE Directive 2002/96/EC
Base segment movable
range
ID input: PS/2 (a ten key pad and a bar code reader)
Data output RS232C
2. IEC 60601-1-2:2001
3. IEC 60825-1:2001 (Laser Class : Class 1 )
4. ISO 15004:1997