A&D HV-G Series Building Set User Manual


 
HV-G Instruction Manual Page 51
16.2 Stream Mode
The scale outputs the current display data at the same time as refreshing the display.
The scale does not output data while in the setting mode.
Averaging function table Refresh rate
f13 0 Normal weighing Approximately 7 times/s while the display is
unstable.
Approximately 4 times/s when the display is stable
f13 1 A persons weight Approximately 4 times/s
f13 1 Animal weighing Approximately 4 times/s
Step 1 Turn off the display.
Press the ON/OFF switch while the ZERO switch is
pressed and held. The function table is displayed.
Step 2 Select a parameter for the baud rate ( f4 )
using the ENTER and ^ switches.
Step 3 Select stream mode ( f5 0 ) of output using
the ENTER and ^ switches.
Step 4 Select format 1 ( f15 0 ) of format using the
ENTER and ^ switches.
Step 5 Press the ENTER switch to store the new
parameters. The scale returns to normal weighing
and outputs the data countinuously.
16.2.1 Preparation and Performance (Examples)