A&D HV-G Series Building Set User Manual


 
HV-G Instruction Manual Page 41
14.2 The Complete Calibration Procedure
14.2.1 Gravity Acceleration Correction
Step1 Turn on the display.
Open the rear cover of the display unit.
Inside the display unit, press and hold the
CAL switch to enter the calibration mode.
Then Cal 0 is displayed.
Step2 Press the ZERO switch to enter the gravity
acceleration correction mode.
Step3 Set your local gravity acceleration using
the following switches.
^ switch Selecting the value of a digit.
< switch Selecting a digit.
Step4 Press the ENTER switch to store the new
value.
Step5 Press the CAL switch again. Then the
scale returns to the normal weighing
mode.
14.2.2 Preparation
Step 6 Confirm the environmental conditions as follows:
Maintain a constant temperature and a stable power supply.
Install the scale on a solid floor or bench where there is no draft,
vibration, strong magnetic fields or direct sunlight.
Consider section “4. Caution”.
Step 7 Display normal weighing for at least 30 minutes to warm up the scale.