A&D HV-G Series Building Set User Manual


 
HV-G Instruction Manual Page 46
Item Display Meaning and purpose
Accumulation When the display becomes a stable value, the scale
accumulates the data automatically. Next accumulation can
perform after the display becomes near-zero.
f8 4 Use: Recording number and mass of articles removed from
the pan. (Put articles on the pan. Press TARE switch at each
removal.)
At each finish of full/dribble batch function, the scale
accumulates the data automatically.
f8 5 Use: Packing articles like a powder, it is used for recording
the bag number and total mass.
The printing condition when using option OP-06 for type V.
f9 0 # No print.
f9 1 Pressing the PRINT switch, the data is output.
Auto-print +
When the display becomes a positive stable value above
f9 2 near-zero, the scale outputs the data automatically. Next
output can be performed after the display becomes near-
zero or negative value.
Print Mode Auto-print +/-
When the display becomes a stable value outside near-zero,
f9 3 the scale outputs the data automatically. Next output can be
performed after the display becomes near-zero.
f9 4 At each finish of full/dribble batch function, the data is output.
f9 5 When the accumulation function is used, the data is output.
The details of the full/dribble batch function (
B$'
)
f10 0 # Reaching final value and pressing the F switch, the current
process is finished.
f10 1 Reaching final value and displaying the stability mark, the
current process is finished automatically.
f10 2 Pressing the SET switch, the scale automatically tares and
starts the full/dribble batch process. Reaching final value and
pressing the F switch, the current process is finished.
Full/Dribble f10 3Pressing the SET switch, the scale automatically tares and
Batch starts the full/dribble batch process. Reaching final value and
sub-function displaying the stability mark, the current process is finished.
Address for RS-422 / RS-485 for option OP-04
f11 00 # Use This address can be set from 01 to 99. Allows a
Address to computer to control a scale with the assigned address.
f11 99 (RS-232C should be set to 00.)
?