A&D HV-G Series Building Set User Manual


 
HV-G Instruction Manual Page 33
Caution
The comparison of the full/dribble batch function is not reversible (One way se-
quence). If the display value becomes less than the final value after the value reached
a predetermined target value, neither HI or LO is on.
The parameters of the upper limit value (HI) and the final value (HI) use the same
memory. The parameters of the lower limit value (LO) and the zero band (LO) use
the same memory.
The upper/lower comparator function, the simple batch function and the full/dribble
batch function cannot be used at the same time because these parameters use
common memory.
Set the zero band greater than the tare value.
Operation
Pressing the SET switch, the scale starts the batch process.
Selecting a parameter from f10 0 or f10 2 of the full/dribble batch sub-function,
the F switch works as the finish switch.
Parameter List and Word Definition
The “gross” is a total weighing value, no tare value is subtracted.
The “net” is a measurement value with the tare value subtracted from the gross.
The “zero band” means the zero detection level.
The “zero point” is the fundamental starting point to weighing.
Comparator
Function table Meaning and purpose
f6 9 Full/Dribble batch function
Full/Dribble batch sub-function
Function table Meaning and purpose
f10 0 Reaching final value and pressing the F switch the current process is
finished.
f10 1 Reaching the final value and displaying the stability mark the current
process is finished automatically.
f10 2 Pressing the SET switch the scale automatically tares and starts the
full/dribble batch process. Reaching final value and pressing the F switch,
the current process is finished.
f10 3 Pressing the SET switch the scale automatically tares and starts the
full/dribble batch process. Reaching the final value and displaying the
stability mark the current process is finished.
Hold
Function table Meaning and purpose
f12 0 The hold function is not used