A&D HV-G Series Building Set User Manual


 
HV-G Instruction Manual Page 18
The “Tare” is used to cancel the mass of a container, receptacle, case, bag, etc.
which is put on the pan to contain the item to be weighed.
Caution
The tare reduces the available weighing range.
The current tare value is reset by pressing the ZERO switch or turning the
scale off. (Reset value is zero.)
Storable preset tare value is within minimum weighing range for the HV-G se-
ries.
7.2.1 Semi-Automatic Tare (Input by Weighing)
Step 1 Put the container onto the pan.
Step 2 Press the TARE switch. The display becomes zero and the net mark
is displayed.
Step 3 It is now possible to put something into the container and to read its
net display.
Step 4 Remove all items from the pan.
7.2.2 Preset Tare (Digital Input of Known Tare)
Step 1 Press and hold the SET switch and press the TARE switch.
Then the blank or stored tare value is displayed. This blank display means
that the tare value is zero (reset value), and blinks.
Step 2 Set the preset tare value by using the following switches.
^ switch selecting the value of the figure.
< switch selecting a figure.
Note: The maximum Preset Tare is equal to the low range maximum capacity.
Step 3 Press the ENTER switch to store the new preset tare value.
The scale displays a net value i.e. the tare value subtracted from the
gross weight value.
Step 3 It is now possible to put something into the container and to read its
net weight.
Step 4 Remove all items from the pan.
7.2 Tare (and Net Display)