A&D HV-G Series Building Set User Manual


 
HV-G Instruction Manual Page 42
14.2.3 Calibration of the Zero Point
Step 8 Press and hold the CAL switch
to enter the calibration mode
after displaying normal
weighing for 30 minutes.
Then the Cal 0 is displayed.
Step 9 With nothing on the pan, press
the ENTER switch while the sta-
ble mark is displayed. The scale
stores the current condition as
the zero point. .
Step 10 The scale displays 5pn1 for
several seconds. If finishing
the calibration mode at this
stage (Zero calibration only),
proceed to step 14.
14.2.4 Span Calibration
Step11 Set the value of the calibration
mass using the following switches.
(This initial value is according to
each product.)
^ switch Selecting the value of
a digit.
< switch Selecting a digit.
Step 12 Place the mass’ selected at
step 11’ onto the pan and press
the ENTER switch while the
stable mark is displayed. The
scale then calculates the span
and stores it.
Step 13 The scale displays end at the
finish.
Remove the mass from the pan.
Step 14 Press the CAL switch to return
to the normal weighing mode.