Function table Meaning & Purpose
f8 4 When the display is a stable value, the scale accumulates the data
automatically. The next accumulation can be performed after the
display returns to near-zero.
Used for recording the number and mass of articles removed from the
pan. (Put the articles on the pan. Press the TARE switch at each
removal.)
f8 5 At each finish of the full/dribble batch function, the scale accumulates
the data automatically.
Used for packaging articles like a powder, it is used for recording the
bag number and total mass.
10.1 Preparation (Setting Parameters)
Step 1 Turn off the display.
Press the ON/OFF switch while the ZERO switch is
pressed and held. The function table is displayed.
Step 2 Press the ENTER switch to display an item of the
accumulation function ( f8 ).
Step 3 Select a parameter of the accumulation function ( f8 1
~ f8 4 ) with the ^ switch.
Step 4 Press the ENTER switch to store the new parameter.
Press the F switch and the ENTER switch to exit from
the function table. Then the scale displays the weighing
mode.