Teledyne API Model 401 O
3
Photometric Calibrator Instruction Manual, 01124, Rev. J2
Table 5-1: Test Channel Output
Test # Name Purpose
1
NONE No output
2
O3 DETECT Outputs O
3
detector
3
REF DETECT Outputs IZS ref. det.
4
SAMP PRESS Outputs sample press.
5
SAMP FLOW Outputs sample flow
6
SAMP TEMP Outputs sample temp.
7
ALAMP TEMP Outputs ana lamp temp.
8
O3 FLOW Outputs O
3
flow
9
O3 LAMP TEMP Outputs O
3
lamp temp
10
CHASSIS TEMP Outputs chassis temp
11
DCPS VOLT Outputs DC power
Table 5-2: Diagnostic Tests
Test # Name Type Purpose
1 D/A OUTPUT Analog Tests analog outputs
5.2.1 D/A Output
This test cycles 3 of the analog output channels from 0% to 100% of Full Scale in 20% FS steps.
It starts by outputting 0 volts to all four channels and displaying a 0% button. Then, every five
seconds, the output is increased 20% FS and the button is changed accordingly. Thus, the button
(and the analog outputs) will cycle through the following values:
0%, 20%, 40%, 60%, 80%, 100%, 0%
To pause the output at the current voltage, press the n% button. To resume automatic cycling,
press the n% button again.
5-2