Leica VT100 A Bouncy Seat User Manual


 
10
Operating Manual V 1.1 – 01/2010
3. Instrument Components and Specifications
3.2 Technical data
General
Approvals: The instrument-specific approval marks are
located on the identification label.
Operating temperature range: +10 °C to +35 °C
Relative humidity: max. 80% non-condensing
Operating temperature range during storage: + 5 °C to +55 °C
Storage humidity: < 80 %
Microtome
Max. section travel length: 0 - 40 mm
Sectioning Speed : 0 - 2.0 mm/s continuously adjustable
Cutting head retraction: c 5-7 mm/s non-adjustable
Blade amplitude: 0 - 3mm, +/- 1.5mm continuously adjustable
Total specimen stroke: 15 mm
Maximum specimen size: 33 x 40 mm
Total angle adjustment: 0 ° to 50 ° continuously adjustable
Tilt, specimen holder: -5 ° - +5 ° in one axis
Electrical data
The instrument is supplied for operation with alternating current in two voltage variants and
for two different frequencies (50 Hz and 60 Hz). For the corresponding product number, refer to
the table on page 14.
Voltage: 110 V AC +/- 15 % 220 V AC +/- 15 %
Max. power consumption: 1 A 0.5 A
Frequency: 50 Hz or 60 Hz (for both voltage variants)
Dimensions and weights
Width: 280 mm
Depth: 460 mm
Height: 310 mm
Weight (without accessories): 15.5 kg