A&D GP-100K Building Set User Manual


 
61
Recalling the memory data
Confirm that the “Data memory (data)” parameter is set to “2”.
1 Press and hold the PRINT key until reCall is displayed, then release the key.
2 Press the PRINT key to enter the memory recall mode. The type of data appears in the upper
left of the display as shown to the right:
Recall the data in memory using the following keys
RE-ZERO key To proceed to the next data set.
MODE key To go back to the previous data set.
PRINT key To transmit the current data using
the RS-232C interface.
With SAMPLE held down, press the CAL key
To delete the data currently displayed.
Note: Deleting the data will not increase the number of data that can be stored.
CAL key To exit the memory recall mode.
3 Press the CAL key to return to the weighing mode.
Transmitting all memory data at one time
Confirm that the “Serial interface (5if)” parameters are set properly. For details, refer to “9.
FUNCTION TABLE” and “15. CONNECTION TO PERIPHERAL EQUIPMENT”.
1 Press and hold the PRINT key until reCall is displayed,
then release the key.
2 Press the SAMPLE key to display out .
3 Press the PRINT key to display Sout no with no” blinking.
5 Press the RE-ZERO key to display out go with “go
blinking.
6 Press the PRINT key to transmit all data using the RS-232C
interface.
7 The balance displays Clear when all data is transmitted.
Press the CAL key to return to the weighing mode.
Deleting all memory data at one time
1 Press and hold the PRINT key until reCall is displayed, then release the key.
2 Press the SAMPLE key several times to display Clear .
3 Press the PRINT key to display SClr no with no” blinking.
4 Press the RE-ZERO key to display SClr go with “go” blinking.
5 Press the PRINT key to delete all data
6 The balance displays end , then reCall when all data is deleted.
7 Press the CAL key to return to the weighing mode.