This indicator means “the balance is measuring calibration data”.
Do not allow vibration or drafts to affect the balance while this
indicator is displayed.
7-2 Automatic Self Calibration (Calibration due to changes in temperature)
This function automatically calibrates the balance when the balance detects an ambient temperature
change. If GLP output is selected in the function table, the balance outputs the calibration report or
stores the data in memory. Automatic self calibration functions even if the display is turned off
(standby state).
Caution
If something is on the weighing pan, the balance judges that it is in use and does not perform
automatic self calibration. To maintain the calibrated state, keep the weighing pan clear while
not in use.
The displays shown below are related to the automatic self calibration.
Indicates that the balance detects a change in ambient
temperature and automatic self calibration will start. If the balance
is not used for a few minutes with this indicator blinking, the
balance performs automatic self calibration. The blinking duration
depends on the environment.
Indicates that the balance is measuring calibration data. Do not
allow vibration or drafts to affect the balance while this indicator is
displayed. After calibration, the balance returns to indicate the
previous display.
Note
The balance can be used while the indicator blinks. But, it is recommended that to maintain
the accuracy, stop using the balance and confirm that there is nothing on the pan and allow
the balance to perform self calibration.