A&D GP-100K Building Set User Manual


 
42
9-5 Description of the Item “Data output mode”
The parameter setting of “Data output mode (prt)” applies to the performance when the “Data
memory (data)” parameter is set to “2” (to store the weighing data) and when the data is transmitted
using the RS-232C interface.
Key mode
When the PRINT key is pressed with the stabilization indictor turned on, the balance outputs or
stores the weighing data and the display blinks one time.
Required setting dout
prt 0 Key mode
Auto print modes A and B
When the displayed value is stable and the conditions of “Auto print polarity”, “Auto print difference”
and reference value are met, the balance outputs or stores the weighing data.
When the PRINT key is pressed with the stabilization indictor turned on, the balance outputs or
stores the data and the display blinks one time.
Mode A: Required setting dout prt 1 Auto print mode A (reference = zero)
dout ap-p Auto print polarity
dout ap-b Auto print difference
Example “For weighing each time a sample is placed and removed,
with “ar-d” set to “1” (to adjust zero after the data is output).”
Mode B: Required setting dout prt 2 Auto print mode B (reference =
last stable value)
dout ap-p Auto print polarity
dout ap-b Auto print difference
Example “For weighing while a sample is added.”
Stream mode
The balance outputs the weighing data continuously regardless of the display condition. The display
does not blink in this mode. This mode is not available and the interval memory mode is used when
the “Data memory (data)” parameter is set to “2” (to store the weighing data).
Required setting dout prt 3 Stream mode
dout data 0 Data memory function is not used.
ba5fnc 5pd Display refresh rate
5if bp5 Baud rate
Example “For monitoring data on a computer”
Caution
The balance may not transmit the data completely at the specified refresh rate, depending on
the baud rate or data added to the weighing data such as time, date and ID number.