INSTALLATION, CONT.
4. With the load cell cable routed into the enclosure, refer to Figure No. 4 and then remove
approximately 5 to 6 inches of the cable outer insulating jacket exposing the internal wires.
5. Next, cut the shield wire so that it extends past the outer jacket approximately 3/4 inch.
6. Now, remove 1/4 inch of insulation from the end of each of the 4 wires (without sense
leads) or 6 wires with sense leads.
1/4 inch
5 to 6 inches
Figure No. 4
3/4 inch
7. Referring to the table below (or on the circuit board) for terminal connections, connect each
wire to terminal block P13. Refer to Figure No. 11 for terminal block location.
LOAD CELL CONNECTOR P13
UPIN NO.U UFunctionU UPIN NO.U UFunctionU
1 + EXCITATION 5 - SIGNAL
2 + SENSE 6 - SENSE
3 + SIGNAL 7 - EXCITATION
8. To terminate a wire, use a small flat blade screwdriver and press down on the release bar
for the terminal. Insert the wire into the opening and remove the screwdriver, allowing the
release bar to return to its original position, locking the wire in place. See Figure No. 5.
Press
Down
Insert
Wire
Figure No. 5
8200-M538-O1 y 225 Installation