Fluke 90100 Child Tracker User Manual


 
Model 90100
Operators Manual
3-6
Pressure Temperature
mm Hg = inch Hg 25.4 K = (5/9) (°F -32) - 273.15
mm Hg = 760 (54,041 - feet ASL) / (54,041 + feet ASL) K = °C + 273.15
mm Hg = 760 (16,472 - meters ASL) / (16,472 + meters ASL)
mm Hg = mB 0.75006
mm Hg = hPa 0.75006
3.5 Basic Operating Equations
This section lists the basic operating equations used within the 35360A TRACKER Display to produce its
dose and dose rate measurement values for display on the VFD. These equations are expressed in
terms of a set of basic operating quantities which are described in Table 3-3.
3.5.1 Charge
Although charge is not a value which can be displayed directly by the 35360A TRACKER Display, it is an
intermediate result upon which the dose value directly depends.
[nC/count] QG ) [counts] ADZ- [counts] (ADC = [nC] Q
×
3.5.2 Current
Although current is not a value which can be displayed directly by the 35360A TRACKER Display, it is an
intermediate result upon which the dose rate value directly depends.
count][(nC/Min)/ IG [counts]) ADZ- [counts] (ADC = [nC/Min] I
×
3.5.3 Absolute Dose
FPGF ADCF Units] Cal Chamber IonUnits/ [Display DUCF
Units/nC] Cal Chamber [Ion ICCF [nC] Q
= Units][Display Display Dose Absolute
×××
×
3.5.4 Absolute Dose Rate
FPGF ADCFUnits] Cal Chamber IonUnits/ [Display DUCF
Units/nC] Cal Chamber Ion [ ICCF [nC/Min] I
= Units/Min][Display Display Rate Absolute
×××
×
3.5.5 Percent-of-Center
Channel Center forDisplay Rate Absolute
Channel Perhiperal forDisplay Rate Absolute
[%] 100 = [%] Channel Perhiperal forDisplay Percentage ×