Adams CFW 60 Building Set User Manual


 
dEC1
Selects the decimal point position
(Options: 0, 0.0, 0.00, 0.000).
C 0.00
InC
Selects the increment size.
(Options: 1, 2, 5 and 10).
inC 10
CAP
Sets the scale capacity. The maximum value
for the scale is as specified for the scale. For
example 60kg.
P2 CAL
CAL
Starts the calibration procedure. See the
details of calibration procedure in section 10.
The sub-parameter “P3 Pro” is used to view the calibration constant used by the
scale, view the A/D counts and set the scale back to default settings. This is
useful if it is not possible to have a good calibration of a scale. It will show the
load cell is causing the A/D to increment correctly through the complete range of
weights needed and you can recall the default settings for the scale. The “trl
value can be modified to cause the scale calibration to change to a known
amount without the need for putting weights on the scale.
trl
Display will show a value that corresponds to
the load cell calibration. If the value is
increased the scale calibration will be
decreased a similar percentage.
CoUnt
Display will show the counts from the A/D
converter. The displayed count should
increase as the weight is added to the load
cell up to and exceeding the capacity of scale.
Typical range for these counts are:
No load 20,000
Maximum 120,000
P3 Pro
rESEt
Scale will reset to factory default settings.
Press [Zero/Enter] to continue or [Print/Esc]
to abort.
Once the value is set, press [Print/Esc] to return to the sub-parameter.
Pressing [Print/Esc] again will return the scale to normal weighing.
@Adam Equipment Company 2006