SETTING OF SERVICE PARAMETERS
SUB-
PARAMETER
FUNCTIONS DESCRIPTION DEFAULT
SETTING
AZn 0
Auto Zero operating range.
(Options: 0.5d, 1.0d, 2d, 4d)
Sets the operational range of the auto
zero function.
1d
0-Auto
Initial Zero range at power on.
(Options: 0%, 2%, 5%, 10% and 20%)
Sets the range for setting a zero when
power is applied. The total capacity of the
scale will not be affected by the new zero
if it is set within 10%. If it is set to a higher
value, the scale may zero but the range
may be limited by load cell, A/D converter
or mechanical stops within the base.
P1 10
(10%)
0-rAnG
Zero range when [Zero] is pressed.
(Options: 4%,10%, 20%, 50%,100%)
Sets the range for setting a zero when
[Zero] is pressed. The total capacity of
the scale will not be affected by new zero
if it is set within 10%. If it is set to a higher
value, the scale may zero but the range
may be limited by load cell, A/D converter
or mechanical stops within the base.
P2 10
(10%)
P1 rEF
SPEEd
Sets the display update speed.
(Options: 7.5, 15, 30 or 60).
SPd 15
The P2 CAL sub-parameter has the basic functions used by the scale to set the
capacity, decimal point position, increment size as well as the calibration
procedures. Do not change the weighing functions as the scales may not
perform as specified. Either the capacity will be wrong, the decimal point will be
in the wrong position or the scale will increment with values that are not as
specified for the scales.
@Adam Equipment Company 2006