Adam Equipment AZplus 30 Building Set User Manual


 
© Adam Equipment Company 2007
16
5.3 CALIBRATION PROCEDURE
The scales are sealed to prevent unauthorised calibration.
To calibrate the scales, access to the PCB inside is needed to place a jumper in K6
position. This is done by removing the tamper-proof label placed on a hole at the bottom of
the scale. Through the hole, pick up the jumper (the jumper will be there placed on one of
the pins on the PCB) and place it in the desired K6 position. If the hole is not provided in
your scale, the security seal has to be broken to remove the top cover and gain access to
the main PCB. Replace the cover and install the screws before starting the calibration.
6.0 SERVICE PARAMETERS
6.1 USING “0000” TO ENTER THE CALIBRATION PARAMETER
After the jumper has been placed in position K6, apply power to the scale.
During the self-test press the [Tare] key.
The display will ask for a code number, “
Pn - - - -“ on the Weight Display.
To carry out the Calibration, enter the number “0000” then press [Tare]. The display will
show “F1” will be enabled.
The display will show “UnloAd” asking you to remove any weight from the scale,
The display will tell you to add weight to the scale by showing “
LoAd” “ 15” “kilo
Add the weight shown, wait for stability then press the [Tare] key.
The display will show “SPAn” “PASS” if the calibration is OK. Or it will show “SPAN” “FAIL” if
calibration could not be completed correctly.
The scale will return to the parameter menu.