3M 490 Doll User Manual


 
4 5
1
SPECIFICATIONS
Power Supply Specifications
Power supply Operating Condition Units
Voltage Range 100‑240 Volts AC
Frequency 50/60 Hertz
Current 0.6 Amps
Environmental Operating Conditions
Environmental Condition Operating Condition Units
Altitude 3000 (max) Meters
Operating Temperature 16 – 40 ºC
Storage Temperature ‑10 to 50 ºC
Operating Relative Humidity 20‑80 (non‑condensing) %
Voltage Range 12 Volts DC
Frequency DC
Current 1.2 Amps
Installation/Over Voltage Category II
Pollution Degree 2
INSTRUCTIONS FOR USE – Initial Set‑up
Choose Location
Place 3M™Attest™ Auto‑reader490 on a firm level surface away from sunlight
and incandescent light. To save counter space, units may be stacked twohigh.
WARNING:
•
Do not place the instrument in environment exposed to sunlight or strong
incandescentlight.
•
Do not place instrument close to any device that emits a strong electro‑magneticfield.
•
Do not use on a vibratingsurface.
Select & Install Correct Power Plug
The 3M™Attest™ Auto‑reader490 power supply comes with multiple blade
configurations. Install the blade configuration appropriate for your country into
the power supply and plug into appropriate AC groundedoutlet.
WARNING: Use only the power supply specified for this product and certified for the
country ofuse.
Wait 30 Minutes Before Using
Once plugged in, allow a 30 minute warm‑up period before placing 3M™Attest™ Super Rapid Readout Biological Indicators into the incubation wells. The C1 caution code will
disappear when proper incubation temperature isreached.
Special Considerations
It is recommended that the unit be left on to eliminate warm‑up periods. The unit may be turned off when it is anticipated that it will not be used for an extendedperiod.