10 11
1
c) History page
Status
History
Well# and Configuration
#8 1492
#3 1491
#7 1492
#6 1492
#2 1491
#1 1491
#2 1491
Start Time
2012-02-22 13:22:58
2012-02-21 10:07:07
2012-02-21 09:51:42
2012-02-21 09:51:18
2012-02-21 09:50:11
2012-02-21 09:49:54
2012-01-2016:02:47
Result
–
+
–
+
–
–
–
3M™ Attest™ Auto-reader 490
ID
1
2
3
4
5
6
7
The 3M™Attest™ Auto‑reader490 stores results for the last 100 incubated BIs on
embedded software. The 3M™Attest™ Auto‑reader490 Web App History page provides
a summary of these last 100results.
d) Configuration page
This page allows the configuration of the 3M™Attest™ Auto‑reader490’s network
settings and Language from the 3M™Attest™ Auto‑reader490 WebApp.
Documenting BI Results using the 3M™Attest™ Auto‑reader490
Web App
Status
History
3M™ Attest™ Auto-reader 490
Copyright © 2010, 2011, 2012 3M, Inc.
ATTEST 1491 ATTEST 1492
1 2 3 4 5
––+
6 7 8 9 10
+–50
SN
100027
2012-02-22 13:33
1. Open the Results Detail page
After inserting a new 3M™Attest™ Super Rapid Readout BI in the 3M™Attest™
Auto‑reader490, it is a best practice to open the 3M™Attest™ Auto‑reader490 Web
App Results Detail page for that BI by clicking on the appropriate well number from the
Statuspage.
2. Enter the specific information for that BI (e.g. BI Lot #, sterilizer/load number,
Technician name, etc.) in the provided editable fields. Use the Notes section to capture
information not prescribed in the provided row titles, for example, you may wish to use
this field to record loaditems.
Copyright © 2010, 2011, 2012 3M, Inc.
3M™ Attest™ Auto-reader 490 Results Detail
Status
History
Print
Result
Start Time
Well# and Configuration
490 Serial Number
Control?
BI Lot#
Load#
Sterilizer#
Technician
Chemical Integrator?
Implant?
Cycle Parameters
Notes
–
2012-02-21 09:51:42
#7 1492
100027
3. The entered information is saved automatically to temporary web browser storage on
the user’s PC every 60 seconds, or by leaving the Results Detailspage.
4. Print/File. Once the BI result is available, print or file the Results Detail page to create
a record. The page may be physically printed and/or saved as an electronicfile.
Important! Data is only stored for the last 100 incubatedBIs.
Embedded software in the 3M™Attest™ Auto‑reader490 saves the data from the
auto‑populated (i.e. green highlighted) fields. The application relies on the temporary
storage in the user’s web browser to save the data in the user‑populated fields. In both
cases, data is only stored for the last 100recordsi.e., the 101st incubated BI will erase
the information saved for the 1st incubated BI. To retain the Results Detail information for
future reference it is best practice to print the page as soon as the BI result isavailable.
Using the BI History Page
Status
History
Well# and Configuration
#8 1492
#3 1491
#7 1492
#6 1492
#2 1491
#1 1491
#2 1491
Start Time
2012-02-22 13:22:58
2012-02-21 10:07:07
2012-02-21 09:51:42
2012-02-21 09:51:18
2012-02-21 09:50:11
2012-02-21 09:49:54
2012-01-2016:02:47
Result
–
+
–
+
–
–
–
3M™ Attest™ Auto-reader 490
ID
1
2
3
4
5
6
7
To view or print an individual record, click on the ID number of the desired record to
navigate to the Results Detail page for thatBI.
Note 1: Column 1, “ID” is the assigned ID number with ID 1 representing the newest
record and ID 100 representing the oldest record. To help find a particular record, all
columns are sortable, simply click once on the column heading to sort in ascending order
and click a second time to reverse theorder.
Note 2: The History page refreshes less frequently than the Status page. Until the BI result
isavailable,theHistorypagedisplays•••intheResultscolumnindicatingincubationis
still in‑process.