e. Are all cable connections to the control module (in the junction box) in place per the
instructions? Are the connectors fully engaged?
f. Are any of the connectors damaged?
g. Are the wires within the connectors attached to the pins? Each wire may have to be
tugged lightly to check if the wire is still crimped to the pin.
4. POST-INSPECTION:
a. If improper or loose connections are found, reconnect them rmly (unless a loose
wire has been found). Plug the power supply AC cord into a live wall socket and
operate the unit through all functions.
b. If the switch seat symbol is rotated or there is evidence of damage to the box and
cable notify Graham-Field so that a switch box and cable can be sent for replacement.
c. If damage to any other electrical cable or component, or disconnected wires are
found, please notify Graham-Field.
5. ACTIONS:
a. If full operation is restored, please advise Graham-Field of your successful outcome.
b. If the fault cannot be corrected by simple re-connection, please inform Graham-Field
of the results of your inspection in detail so that we may determine the appropriate
action to be taken.
If you have any questions about the setup and / or operation of the Heat and Massage
Unit, please call the Graham-Field Technical Support Group at 770-447-1609.