TAPI Model 265A Chemiluminescence Ozone Analyzer
Manual Addendum, Rev. B, DCN 5254
23
PRINTED DOCUMENTS ARE UNCONTROLLED
5 MAINTENANCE
5.1 Maintenance Schedule
NOTE
The operations outlined in this chapter are to be
performed by qualified maintenance personnel only.
Table 5-1: Preventative Maintenance Schedule
Item Maintenance Interval Reference Section
TEST functions Review monthly See M200A manual, Table 8-1
Zero/Span calibration Annually or after repairs See M200A manual, Table 8-1
Zero/Span checks Daily See M200A manual, Table 8-1
Particulate filter Weekly as needed See M200A manual, Table 8-1
Reaction cell window Clean annually or as necessary See M200A manual, Table 8-1
Nitric oxide flow Check TEST function every
month
Section 1-3
See also M200A manual, Table
8-1 (O
3
flows)
Sample Flow Check TEST function every
month
See M200A manual, Table 8-1
Pneumatic lines Examine every 12 months,
clean if necessary.
See M200A manual, Table 8-1
Factory calibration Calibrate each year or after
repairs
See M200A manual, Table 8-1
Leak Check Check every year or after other
maintenance.
See M200A manual, Table 8-1
Reaction cell O-rings Replace every 12 months See M200A manual, Table 8-1
Other O-rings Replace every 12 months See M200A manual, Table 8-1