TAPI Model 265A Chemiluminescence Ozone Analyzer
Manual Addendum, Rev. B, DCN 5254
13
PRINTED DOCUMENTS ARE UNCONTROLLED
2.3 Test and Calibration Values
Table 2-2: Final Test and Calibration Values
TEST Values
Observed
Value
Units Nominal Range Reference Section
RANGE ppb 10-20000
NOISE ppb 0.0 – 7.0
SAMP FLW cm³/min 500 ± 50
NO FLOW cm³/min 5 ± 1
PMT mV 0-5000
AUTOZERO mV -10 to +50
HVPS V 400 – 850 constant
DCPS mV 2500 ± 200
RCELL TEMP °C 50 ± 2
BOX TEMP °C 8-40
PMT TEMP °C 8 ± 2
RCEL PRES IN-Hg-A 2 - 10 constant
SAMP PRES IN-Hg-A 20 - 30 constant
NO PRES PSIG 3 - 20 constant
Electric Test &Optic Test
Electric Test
PMT Volts mV 2000 ± 200
O
3
Conc ppb 1000 ± 100
OPTIC TEST
PMT Volts mV 2000 ± 200
O
3
Conc ppb 1000 ± 100
(table continued)