S12Q-GOLD & S24Q-GOLD TROUBLESHOOTING GUIDE
SYSTEM STATUS
REMARKS
UV
INTENSITY
(LED)
AUDIBLE
ALARM
UV LAMP
GREEN OFF ON
Correct operating conditions unit is functioning properly.
YELLOW OFF ON
System performance has declined due to stained or dirty
quartz sleeve and or sensor lens, or the lamp is beyond its
useful life.
Clean sleeve and sensor probe lens or replace if required.
Replace lamp with manufacturer replacement.
Replace sensor with manufacturer replacement.
Always Wet Sensor Lens prior to inserting sensor.
check water quality (see page 2 for water parameters)
FLASHING
RED
INTERMIT-
TENT
ON
System failure is imminent due to stained or dirty quartz
sleeve or lamp is beyond useful life
Clean sleeve and sensor probe lens or replace if required.
Replace lamp with manufacturer replacement.
Replace sensor with manufacturer replacement.
Always W
et Sensor Lens prior to inserting sensor.
check water quality (see page 2 for water parameters)
RED ON ON
UV Sensor has not detected enough UV Intensity to
adequately protect the water due to stained or dirty
quartz sleeve or the lamp is beyond its useful life.
Clean sleeve and sensor probe lens or replace if required.
Replace lamp with manufacturer replacement.
Replace sensor with manufacturer replacement.
Always Wet Sensor Lens prior to inserting sensor.
check water quality (see page 2 for water parameters)
RED ON OFF
Ballast not functioning,
Reset Ballast by removing power for 30 seconds
then reapply.
Check wiring to ensure ballast is plugged into
circuit board.
Replace ballast with manufacturer replacement
Lamp is spent. Replace lamp with manufacturer
replacement
GREEN OFF OFF
Defective UV sensor assembly. Replace sensor assembly with
manufacturer replacement.
Calibration board improperly seated. Check board to ensure
proper engagement with main circuit board.
Defective sensor assembly.