Adams AQT SERIES Building Set User Manual


 
8.2 AUTO CALIBRATION FOR ALL OTHER MODELS
Before entering calibration mode, press [On/Off] to turn the power
on for 1 minute.
Enter calibration mode
Press [On/Off] to turn the power off.
Press [Unit] first and then press [On/Off], while keeping the [Unit]
key pressed. The display will show the last calibration value used
(For example, “ L – 1000”), press [Tare] to choose between the two
calibration weight values. When you have made your selection press
[Unit] and the display will show the A/D value.
Calibration
Wait for the stable indicator to be displayed,
Press [Unit] -the display will show "00SAVE". After 2 seconds, the
display will return to A/D value.
Place the required weight on the platform. This is the weight that you
selected as your calibration weight value earlier.
After the stable indicator is displayed press [Unit]
The display will show "00SAVE". The calibration in complete now.
Press [On/Off] to turn the power off. Press [On/Off] to turn on the
power again,
Place full weight (eg.2000g for AQT2000) on the platform to verify
weighing is correct. If not, repeat the calibration procedure
.
9.0 ERROR MESSAGES
EEEEHI Overload
EEEELO A/D value too low
Low battery
Note: Calibration may be required before weighing. Read section 8.0 on
"CALIBRATION" first and if necessary calibrate your balance for accurate
weighing.