8.0 CALIBRATION
Calibration may be required when the balance is initially installed or
if the balance is moved from a location.
With time and use, mechanical deviations may occur. Therefore
calibrate the balance whenever the best accuracy is required or
when any deviation is observed during weighing a test weight.
Before calibrating it is necessary to have the proper calibration
weights. The weights should be known to an accuracy that is
appropriate for the balance being calibrated, for example, OIML
Class M1 type or ASTM E617 Class 4. If you do not have proper
weights do not attempt calibration.
Model AQT-200 AQT- 250 AQT-600 AQT-1500 AQT-2600 AQT-2000 AQT-5000
Calibration
Weight
100g
Or
200g
200g
Or
250g
500g
Or
600g
1000g
Or
1500g
1000g
Or
2000g
1000g
Or
2000g
2000g
Or
5000g
8.1 AUTO CALIBRATION (FOR AQT 200 ONLY)
Before entering calibration mode, press [On/Off] to turn the power
on for 1 minute.
Enter calibration mode
Press [On/Off] to turn the power off.
Press [Unit] first and then press [On/Off], while keeping the [Unit]
key pressed. The display will show the A/D value. Then press [Unit]
and the display will show the last calibration value used, press [Tare]
to choose either 100g or 200g.
Calibration
Wait for the stable indicator to be displayed,
Press [Unit] - the display will show “L 100" or “L 200”, with the "L"
flashing.
Place the appropriate weight on the platform. A beep will be heard
after a few seconds and the display will show the weights value.
Press [On/Off], to turn off the power.
Press [On/Off] to turn on the power again. Place maximum weight
(200g) on the platform to verify weighing is correct. If not, repeat the
calibration.