Adam Equipment GFK 150M Building Set User Manual


 
49
| P a g e © Adam Equipment Company 2010
14. ERROR MESSAGES
During the initial power-on testing or during operation, the scale may show an error
message. The meaning of the error messages is described below.
If an error message is shown, repeat the step that caused the message. If the error
message is still shown then contact your dealer for support.
ERROR
CODE
DESCRIPTION POSSIBLE CAUSES
Err 1
Time input Error
Invalid time entry such as “268970”
for the time format “H-m-S”.
Err 2
Date input Error
34th day of a month is an invalid
entry.
Err 4
Initial Zero is greater than
allowed (4% of maximum
capacity) when power is
turned on or when the
[Zero/Enter] key is pressed.
Weight on the platform when turning
the scale on.
Excessive weight on the platform
when zeroing the scale.
Platform is not installed.
Improper calibration of the scale.
Damaged load cell.
Damaged Electronics.
Err 6
A/D count is not correct
when turning the scale on.
Load cell is damaged.
Electronics is damaged.
Err 7
Percent input error
Percent function is entered with no
reference mass on the platform.
Err 8
High limit input error
Low limit is set first, then the high
limit is set lower than the low limit
and high limit not equal to zero.