3M 1210 Musical Toy Instrument User Manual


 
Power “ON” Display and Startup
The display of a switched “ON” pipettor indicates that programming has been reset
to default settings.
1. Switch the unit “ON”. Display:
2. Press the Start Button twice. The display
shows that the unit is ready for the
default 1:5 dilution (Mode d1, 5mL) or
for program changes.
Switching the pipettor “OFF” and then
“ON” will reset the unit to the default
settings (Mode d1, 1:5, 5mL dilution).
Mode Selection and Mode Recall
Selecting available modes on the pipettors is done with the use of these three keys:
1. Press once to display the default dilution mode of the pipettor.
2. Press repeatedly to view each of the modes available.
3. Press when the desired mode is displayed. This will enter the mode
selection and allow you to use this mode.
Mode recall allows you to flip back and forth between the last two
modes used.
or may not be used in the middle of an aspiration or
dispensing cycle.
Pipettor Programming
8
Pipettor Programming
7
The Program Menu
1. The following five dilutions have been pre-programmed.
Dilution Diluent Sample Total
Mode Ratio Volume Volume Volume
d1 1:5 4.00mL 1.00mL 5.00mL
d2 1:10 4.50mL 0.50mL 5.00mL
d3 1:20 4.75mL 0.25mL 5.00mL
d4 1:5 0.80mL 0.20mL 1.00mL
d5 1:10 0.90mL 0.10mL 1.00mL
2. Repeat Dispense 1mL Mode-
This mode performs repeat 1mL dispensing.
Mode Volume Repetitions
rd 1.00mL 1-5
3. Pipette Modes -
These modes perform standard 1mL and 5mL pipetting with blow-out.
Mode Volume
P1 1.00mL
P5 5.00mL
4. Open Dilution Mode -
This mode allows selection of any dilution ratio within the range of the pipettor
Diluent Sample Max.
Volume Volume Total
Mode Increment Range Range Volume
d0 0.05mL 0.10-5.00mL 0.00-4.90mL. 5.00mL
Page 4 of 15 - June 2001