Pipettor Programming
20
Speed Selection
Speed changes are made using these four keys:
Speed "1" is slowest and Speed "5" is fastest.
1. Press to display the
aspiration speed.
2. Press or to select a
new aspiration speed.
3. Press when the desired
aspiration speed is displayed.
Dispensing speed is then displayed.
4. Press or to select a
new dispensing speed.
5. Press when the desired
dispensing speed is displayed. The
display then returns to the diluent
aspiration readout.
may not be used in the middle
of any aspiration or dispensing cycle.
Pipettor Programming
19
7. Attach a pipette tip to the tip cone.
The pipettor is now ready to begin the
dilution. The display indicates that the
pipettor is ready to aspirate the selected
diluent volume.
8. Place the tip in the diluent and press the
Start Button. The unit will aspirate the
diluent volume selected.
The following display indicates that the
unit is ready to aspirate the selected
sample volume.
9. Place the tip in the sample solution
and press the Start Button to aspirate
0.50mL.
The next display indicates that the unit
is ready to dispense the total volume
with blowout.
10. Position the pipette tip to dispense and
press the Start Button. The unit will
dispense the full amount from the tip.
The display now indicates that the
unit is ready to repeat the dilution
sequence, beginning with the diluent
volume aspiration.
Page 10 of 15 - June 2001