3M 247 Crib Toy User Manual


 
3M
TM
Ranger
TM
Irrigation Fluid Warming System
English
13
202456B English
Check the 3M
TM
Ranger
TM
system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202456B

1. Unplug the irrigation uid warming unit.
2. Unfold the cleaning tool. Wet the foam pads
with a nonabrasive solution, such as Alconox
TM
brand detergent.
3. Insert the tool from the back of the unit and
pull the tool all the way out from the front.
4. Rinse the tool with water and repeat 3 times. Discard the tool according to institutional protocol.
5. Wipe o the unit to remove excess uid.
  , - :
1. Spray a nonabrasive solution inside the slot of the warming unit and let sit for 15-20 minutes.
2. Clean the unit by using the cleaning tool.
NOTE: You may use a nonmetal instrument, such as a cotton swab, to clean the upper channels. If you are
unable to adequately clean the unit, call 3M Patient Warming technical service.
Storage
Store all components in a cool, dry place when not in use.
Servicing
ere are no user-serviceable parts in the Ranger irrigation uid warming unit. All service must be performed by
3M Patient Warming or an authorized service technician.
Call 3M Patient Warming technical service at 800-733-7775 or 952-947-1200 for service information.
Alconox is a trademark of Alconox, Inc.