Check the 3M
TM
Ranger
TM
system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202456B
WARNING:
1. To reduce the risks associated with hazardous voltage and re and thermal energy hazards:
• Do not substitute other devices (i.e. Model 245) for the Ranger irrigation uid warming unit or Ranger
irrigation disposable sets.
• Do not continue use of the unit if the over-temperature alert continues to sound and the temperature
does not return to the set point temperature. Immediately stop uid ow and discard the disposable
set. Have the warming unit tested by a biomedical technician or call 3M Patient Warming technical
service.
2. To reduce the risks associated with hazardous voltage and re:
• Do not modify or service this device, and do not open the warming unit case- there are no user-
serviceable parts in the Ranger warming unit.
• Connect power cord to receptacles marked “Hospital Only,” “Hospital Grade,” or a reliably grounded
outlet.
• Use only the power cord specied for this product and certied for the country of use.
• Do not allow the power cord to get wet.
• Do not use the Ranger warming system when it appears the warming unit, power cord, or disposable
is damaged. Use only 3M specied replacement parts.
• Keep power cord visible and accessible at all times. e plug on the power cord serves as the
disconnect device. e wall socket outlet shall be as close as practical and shall be easily accessible.
3. To reduce the risks associated with air embolism:
• Never use uids if air bubbles are present in the uid line.
CAUTION:
1. Not for I.V. use.
2. To reduce the risks associated with cross-contamination:
• e cleaning tool provides only supercial cleaning, it does not disinfect or sterilize the interior of the
unit.
3. To reduce the risks associated with impact and facility medical device damage:
• Clamp the Ranger irrigation uid warming unit to an I.V. pole with a minimum 14” (35.6 cm) radius
wheelbase and at a height no higher than 44” (112 cm).
4. To reduce the risks associated with environmental contamination:
• Follow applicable regulations when disposing of this device or any of its electronic components.
NOTICE:
1. Federal law (USA) restricts this device to sale by or on the order of a licensed healthcare professional.
2. To avoid device damage:
• Do not clean the warming unit with solvents. Damage to the case, label, and internal components may
result.
• Do not immerse the warming unit in cleaning or sterilizing solutions. e unit is not liquid proof.
• Do not insert metallic instruments in the warming unit.
• Do not use abrasive materials or solutions to clean the heater plates.
• Do not allow spills to dry inside the unit, as this may make it more dicult to clean the unit.
3. e Ranger irrigation uid warming unit meets medical electronic interference requirements. If radio
frequency interference with other equipment should occur, connect the unit to a dierent power source.