Irrigation Fluid Warming System
English
6
Check the 3M
TM
Ranger
TM
system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202456B
WARNING:
1. To reduce the risks associated with hazardous voltage and re and thermal energy hazards:
• Do not substitute other devices (i.e. Model 245) for the Ranger irrigation uid warming unit or Ranger
irrigation disposable sets.
• Do not continue use of the unit if the over-temperature alert continues to sound and the temperature
does not return to the set point temperature. Immediately stop uid ow and discard the disposable
set. Have the warming unit tested by a biomedical technician or call 3M Patient Warming technical
service.
2. To reduce the risks associated with hazardous voltage and re:
• Do not modify or service this device, and do not open the warming unit case- there are no user-
serviceable parts in the Ranger warming unit.
• Connect power cord to receptacles marked “Hospital Only,” “Hospital Grade,” or a reliably grounded
outlet.
• Use only the power cord specied for this product and certied for the country of use.
• Do not allow the power cord to get wet.
• Do not use the Ranger warming system when it appears the warming unit, power cord, or disposable
is damaged. Use only 3M specied replacement parts.
• Keep power cord visible and accessible at all times. e plug on the power cord serves as the
disconnect device. e wall socket outlet shall be as close as practical and shall be easily accessible.
3. To reduce the risks associated with air embolism:
• Never use uids if air bubbles are present in the uid line.
CAUTION:
1. Not for I.V. use.
2. To reduce the risks associated with cross-contamination:
• e cleaning tool provides only supercial cleaning, it does not disinfect or sterilize the interior of the
unit.
3. To reduce the risks associated with impact and facility medical device damage:
• Clamp the Ranger irrigation uid warming unit to an I.V. pole with a minimum 14” (35.6 cm) radius
wheelbase and at a height no higher than 44” (112 cm).
4. To reduce the risks associated with environmental contamination:
• Follow applicable regulations when disposing of this device or any of its electronic components.
NOTICE:
1. Federal law (USA) restricts this device to sale by or on the order of a licensed healthcare professional.
2. To avoid device damage:
• Do not clean the warming unit with solvents. Damage to the case, label, and internal components may
result.
• Do not immerse the warming unit in cleaning or sterilizing solutions. e unit is not liquid proof.
• Do not insert metallic instruments in the warming unit.
• Do not use abrasive materials or solutions to clean the heater plates.
• Do not allow spills to dry inside the unit, as this may make it more dicult to clean the unit.
3. e Ranger irrigation uid warming unit meets medical electronic interference requirements. If radio
frequency interference with other equipment should occur, connect the unit to a dierent power source.