A SERVICE OF

logo

To set up the intermodulation distortion test
Performance Test only
This test verifies that the Agilent 35670A meets its FFT dynamic range
specification for intermodulation distortion. In this test, two signals are combined
to provide a composite signal to all channels. The intermodulation products are
found at the sum (F1 + F2) and difference (F1 2F2) frequencies. The analyzer
measures the amplitude of each intermodulation product.
1 2ch Using an HP 3326A
1 4ch Using an HP 3326A
Verifying Specifications Agilent 35670A
To set up the intermodulation distortion test
3-28