While monitoring the Y: value, adjust A2 R543 in the lower slot for a
minimum marker value.
Disconnect the BNC cable from the analyzer’s CH 3 connector and connect
to the CH 4 connector.
Press the the following keys:
[
Rtn ]
[
CHANNEL 4 ADJUSTMNT ]
[
CMRR ]
Wait for the analyzer to set up the adjustment. The analyzer is ready when
the adjustment message appears on the screen.
While monitoring the Y: value, adjust A2 R543 in the upper slot for a
minimum marker value.
Set the power switch to off ( O ) and disconnect the minigrabber clips from
A5 TP8.
Adjusting the Analyzer Agilent 35670A
To adjust common mode rejection
5-16