14423-INS-LAB-RevE13 7
WARNING:UseonlyprobesprovidedbyGFHealth
Products,Inc.intendedtobeusedwiththeGrafco
PocketDoppler.
WARNING:Donotstretchtheprobecablefurther
than6.5feet(2meters).
WARNING:ElectromagneticInterference:Before
operation,ensurethattheenvironmentinwhichthe
deviceisoperatedisnotsubjecttoanysourceof
strongelectromagneticinterference,suchasradio
transmitters,mobiletelephones,etc.
WARNING:Thefollowingsafetychecks/testsmust
beperformedatleastonceeverytwoyearsor
asspeciedinyourfacility'stestandinspection
protocol,byaqualiedpersonwithadequate
training,knowledge,andpracticalexperience:
• Inspecttheequipmentformechanicaland
functionaldamage.
• Ensurethesafetylabelsarelegible.
• Verifythatthedevicefunctionsproperlyas
describedinthismanual.
• Testaccordingtothepregnantwoman'sleakage
current(IEC60601-1/1988:Limit:100uA(B)).
Theleakagecurrentshouldneverexceedthe
speciedlimit.Recordthedataandstorewith
theGrafcoPocketDoppler.Ifthedeviceisnot
functioningproperlyorfailsanyoftheabovetests,
contactGFHealthProducts,Inc.TechnicalSupport
atthephonenumberonthebackcover,oryourGF
HealthProducts,Inc.authorizeddistributor.