Info 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Info 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reection from
structures, objects and people.
a: Field strengths from xed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio,
AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment
due to xed RF transmitters, an electromagnetic site survey should be
considered. If the measured eld strength in the location in which the
Grafco Ultrasound Pocket Doppler is used exceeds the applicable RF
compliance level above, the Grafco Ultrasound Pocket Doppler should
be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or
relocating the Grafco Ultrasound Pocket Doppler.
b: Over the frequency range 150 kHz to 80 MHz, eld strengths should
be less than 3 V/m.




The Grafco Ultrasound Pocket Doppler is intended for use in an
electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the device can help prevent
electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the
Grafco Ultrasound Pocket Doppler as recommended below, according to the
maximum output power of the communications equipment.









0.01 0.1167 0.2334
0.1 0.3689 0.7378
1 1.1667 2.3334
10 3.6893 7.3786
100 11.6667 23.3334