A&D GP-60K Building Set User Manual


 
32
7-6 Correcting the Internal Mass Value
The GP series balance can correct the internal mass value within the range shown below. This
function corrects the internal mass value to conform to an external weight. The corrected mass value
is maintained in non-volatile memory even if the AC adapter is removed.
The internal mass value is corrected as follows:
Model Correction reference value Correction range
GP-12K 10 kg
GP-20K
GP-30K/GP-30KS
GP-32K/ GP-32KS
GP-61K/ GP-61KS
20 kg
GP-40K 40 kg
-5.0 g to +5.0 g
GP-60K/ GP-60KS 60 kg -50 g to +50 g
GP-100K/GP-100KS
GP-102K
100 kg -50 g to +50 g
Example: Using the GP-20K