12. PROGRAMMABLE-UNIT
This is a programmable unit conversion function. It multiplies the weighing data in kilograms by an
arbitrary coefficient set in the function table and displays the result.
The coefficient must be within the range between the minimum and maximum shown below. If the
coefficient set is beyond the range, an error is displayed and the balance returns to the coefficient
setting mode, prompting to enter an appropriate value. A coefficient of 1 was set at the factory.
Model Minimum coefficient Maximum coefficient
GP-12K/20K/30K/30KS/40K/61K/61KS 10000
GP-60K/60KS/100K/100KS 1000
GP-32K/32KS 1000
GP-102K
0.000001
100
Operation
1 Press and hold the SAMPLE key until ba5fnc of the function table is displayed, then release the
key.
2 Press the SAMPLE key several times to display mltCadj .
3 Press the PRINT key. The balance enters the mode to confirm or set the coefficient.
Confirming the coefficient
4 The current coefficient is displayed with the first digit blinking.
When it is not to be changed, press the CAL key and proceed to step 6.
When it is to be changed, press the RE-ZERO key
and proceed to step 5.
Setting the coefficient
5 Set the coefficient using the following keys.
SAMPLE key To select a digit to change the
value. The selected digit blinks.
RE-ZERO key To change the value.
MODE key To change the decimal point
position.
Each time the switch is pressed,
the decimal point position
changes as follows:
PRINT key To store the new setting, display end and go to step 6.
CAL key To cancel the new setting and go to step 6.
Quitting the operation
6 The balance displays Unit . Press the CAL key to exit the programmable-unit function and
return to the weighing mode.
Using the function
Press the MODE key to select the programmable-unit (no display on the unit section). Perform
weighing as described in “5-1 Basic Operation (Kilogram Mode)”. After weighing, the balance
displays the result (weighing data in kilograms x coefficient).