A&D GP-32KS Building Set User Manual


 
45
9-7 Description of the Data Format Added to the Weighing Data
ID number dout 5-id 1
The number to identify a specific balance.
This format consists of seven characters excluding the terminator.
Data number dout d-no 1
This format outputs the data number just before the data is transmitted using the RS-232C interface.
This format consists of six characters excluding the terminator.
When CSV format (5if type 5) is selected, the period (
. ) is replaced with a comma ( , ).
Date dout 5-td 2 or 3
The date output order can be changed in “Clock (Cl adj)”. Outputs the year in four-digit
format.
Time dout 5-td 1 or 3
Outputs time in 24-hour format.
Tare value
When the tare value in memory is recalled, the tare value is output before the weighing data.
Tare value
Weighing data (Net value)
Comparison results
By setting “Comparison results (Cp-r)” of the function table to “1”, the comparison results can be
added to the data output using the RS-232C serial interface. Use A&D standard format (type 0).
The comparison results are added after the header in A&D standard format as below.
Note
When the data described above is added to the weighing data, the output is in the
following order: ID number, Data number, Date, Time and Weighing data.