53
Setting the function table
2 Press and hold the SAMPLE key until ba5fnc of the function table is displayed, then release the key.
3 Press the SAMPLE key several times to display Cp fnc .
4 Press the PRINT key.
5 Press the SAMPLE key several times to display Cp-b 0 .
6 Press the RE-ZERO key to display Cp-b 1 .
Note
To disable the main display comparison funciton, set the “Main display comparison (Cp-b)”
parameter to “0”.
7 Press the PRINT key to store the setting.
8 Press the CAL key to return to the weighing mode.
Setting the comparator values
Setting the comparator values as described in
the previous section.
Setting example (Cp 3=Continuous comparison,
excluding “near zero”)
Using the main display comparison function
1 Press the RE-ZERO key to set the
display to zero.
2 Place a sample on the pan. The
balance performs a comparison using
the specified comparison values and
displays the comparison results, HI,
OK or LO.
3 Each time the MODE key is pressed,
the balance switches between the
standard display and the main display
comparison. Note that “g0” appears
for OK.
Notes
While the main display comparison
function is in use, the processing indicator
illuminates as shown in the illustration.
If the comparison is not performed, for example, because the weight value is near zero or
unstable, the balance displays the weight value even when the main display comparison
function is used.
Even while the main display comparison function is in use, the balance re-zeroing and data
output is possible.
Only the unit selected before this function can be used.
While the main display comparison function is in use, the data memory function is not
available.