5.0 WEIGHING
5.1
BEFORE
WEIGHING
Check the setting of the auto shut off function, see section 6.2. If the auto
shut off function is on, the scale will shut off after a period of time if the scale
is not used and it has a stable reading.
5.2
WEIGHING
Press [On/Off] to turn on the scale. The display will show the software
revision then show the battery voltage “X.X vol” .Then it will begin it’s
self-checking test from 0 to 9. When reading is stable, the display will
show "0.00", stable symbol, zero symbol, and the weighing unit used
the last time the scale was used.
Press [Units/
] to select units. Press [Units/
] to select units from
"g" / "Kg" / "lb" / "oz"/”CUS”. Only the weighing units enabled will be
displayed, See section 6.9.
Check if the reading is “0.00”, if not, press [Zero/
ESC
]. “0.00” and the
zero symbol will be displayed.
Put the item to be weighed on the pan. When the reading is stable, the
net weight will be shown.
If a container is used, put the container on the pan and press the [Tare/
] key when
the reading is stable. The display will show “0.00” and the zero symbol. If the
container is more than 2% of the capacity, the NET symbol will be on. Load the
container or remove material from the container, the scale will show the net weight
of the material added or removed.