Teledyne M200EU Card Game User Manual


 
Ultra Sensitivity Model M200EU NO/NO
2
/NOx Analyzer Addendum to M200E
2 TAPI 05385 Rev B
PRINTED DOCUMENTS ARE UNCONTROLLED
2. SPECIFICATIONS, AGENCY
APPROVALS, AND WARRANTY
Min/Max Range
(Physical Analog Output)
Min: 0-5 ppb of Full Scale
Max 0-2000 ppb of Full Scale
Measurement Units ppb, ug/m
3
(user selectable Single, independent or auto ranges)
Noise at Zero
1
<25 ppt RMS
Noise at Span
1
<0.25% RMS of reading above 50 ppb
Lower Detectable Limit
1
50 ppt RMS
Zero Drift
2
<0.1 ppb / 24 hours
<0.2 ppb / 7 days
Span Drift
2
<0.5% of reading or 50 ppt RMS whichever is greater / 24 hrs
<1% of reading or 100 ppt RMS whichever is greater / 7 days
Linearity
1% of full scale or 0.1 ppb whichever is greater
Precision 0.5% of reading above 50 ppb
Lag Time
1
20 sec
Rise / Fall Time
1, 3
95% in <50 sec
Sample Flow Rate 1000 cc/min. ± 10%
Temperature Range
20-30
o
C within drift and noise specifications, and operating with EPA equivalency
5-35
o
C safe operating range
Humidity 0-95% RH non-condensing
Hydrocarbon Interference
Propylene rejection ratio > 20,000:1
Ethylene rejection ratio > 40,000:1
Voltage Coefficient < 0.1% per V
Dimensions H x W x D 7" x 17" x 23.5" (178mm x 432 mm x 597 mm)
Weight, Analyzer 18 kg (40 lbs)
Weight, Ext Pump Pack 9.5 kg (21 lbs)
Power, Analyzer 100V~ 50/60 Hz, 120V~ 60 Hz, 230V~ 50 Hz, 2.5A, 125 watts
Power, Analyzer
4
230V~ 50 Hz, 2.5A
Power, Ext Pump 110V~ 60 Hz, 220V~ 50 Hz, 240V~ 50Hz, 150 watts
Power, Ext Pump
4
230V~ 50 Hz, 2.2A
Analog Outputs Four (4) Outputs
Analog Output Ranges
All outputs: 100 mV, 1, 5, 10V
Three outputs convertible to 4-20 mA isolated current loop
All ranges with 5% under/over range
Analog Output Resolution 1 part in 4096 of selected full-scale voltage
Status Outputs 8 Status Outputs from opto-isolators, 7 defined, 1 spare
Control Inputs 6 Control Inputs, 4 defined, 2 spare