Pride Mobility Quantum 614 3MP High Chair User Manual


 
Quantum 614 Series www.quantumrehab.com 17
II. SAFETY
WARNING! Pride strongly recommends that you do not smoke cigarettes while seated in your
power chair, although the power chair has passed the necessary testing requirements for cigarette
smoking. You must adhere to the following safety guidelines if you decide to smoke cigarettes
while seated in your power chair.
Do not leave lit cigarettes unattended.
Keep ashtrays a safe distance from the seat cushions.
Always make sure cigarettes are completely extinguished before disposal.
Removable Parts
WARNING! Do not attempt to lift or move a power chair by any of its removable parts, including the
armrests, seat, front riggings, controller, and shrouds.
Preventing Unintended Movement
WARNING! If you anticipate being seated in a stationary position for an extended period of time,
turn off the power. This will prevent unexpected motion from inadvertent joystick contact. This
will also eliminate the possibility of unintended chair movement from electromagnetic (EM) sources.
Electromagnetic and Radio Frequency Interference (EMI/RFI)
WARNING! Laboratory tests have shown that electromagnetic and radio frequency waves can
have an adverse affect on the performance of electrically-powered mobility vehicles.
Electromagnetic and Radio Frequency Interference can come from sources such as cellular phones, mobile two-way
radios (such as walkie-talkies), radio stations, TV stations, amateur radio (HAM) transmitters, wireless computer links,
microwave signals, paging transmitters, and medium-range mobile transceivers used by emergency vehicles. In some
cases, these waves can cause unintended movement or damage to the control system. Every electrically-powered mobility
vehicle has an immunity (or resistance) to EMI. The higher the immunity level, the greater the protection against EMI. This
product has been tested and has passed at an immunity level of 20 V/m.
WARNING! Be aware that cell phones, two-way radios, laptops, and other types of radio transmitters
may cause unintended movement of your electrically-powered mobility vehicle due to EMI.
Exercise caution when using any of these items while operating your mobility vehicle and avoid
coming into close proximity of radio and TV stations.
WARNING! The addition of accessories or components to the electrically-powered mobility vehicle
can increase the susceptibility of the vehicle to EMI. Do not modify your power chair in any way
not authorized by Pride.
WARNING! The electrically-powered mobility vehicle itself can disturb the performance of other
electrical devices located nearby, such as alarm systems.
NOTE: For further information on EMI/RFI, go to the Resource Center on www.pridemobility.com. If
unintended motion or brake release occurs, turn your power chair off as soon as it is safe to do so. Call Pride
at 800-424-8205 to report the incident.