Specifications
Monitor (M2922A)
Federal Law restricts this device to use by or on the
order of a physician.
1
Standards Compliance
The FM-2 Portable Antepartum Fetal Monitor
complies with the requirements of the Medical
Device Directive 93/42/EEC and the following
safety standards:
EN60601-1/IEC60601-1, EN60601-1-1,
EN60101-2-37
IEC 1157
IEC 60601-1
UL2601-1
CAN/CSA C22.2#601.1-M90
Class II equipment, double insulated
Continuous operation
Electromagnetic Compatibility
Complies with EN60601-1-2
Environmental
Operating Temp: -10
o
C to 45
o
C
(50
o
F to 110
o
F)
Storage Temp: -20
o
C to 60
o
C
(-29
o
F to 175
o
F)
Relative Humidity: 20% to 90% non
condensing
Altitude: 0-3048m (10,000 ft)
Dimensions
Height: 15.24 cm (6.0 inches)
Width: 26.4 cm (10.4 inches)
Depth: 13.7 cm (5.4 inches)
Weight: 2.27 kg (5lb)
Power - Battery
Recharge without Maximum 8 hours to
monitoring: full recharge when no
monitoring occurs (typical
4.75 hours).
Low power warning: Approximately 30 minutes
at 25
o
C.
Power - External Power Supply
AC-powered 110-120V~, 50-60Hz,
0.15A
or
220-240V~, 50-60Hz,
0.08A
Display
Backlight: Cold cathode fluorescent
Active area: 3.78 x 2.83
(320 x 240 dots)
Resolution: 85 dots per inch
Contrast: 10:1
Ultrasound
I
sata
: < 4mW
BPM range:
Paper scale: USA paper: 30-240 BPM,
International Paper: 50-210
BPM.
Ultrasound
processing range: 50-240 BPM.
Accuracy:
±1% ±1 BPM.
Leakage current:
10 µ A @ 264 VAC applied
to transducer.
Isolation: >4 kV RMS, Type BF
applied part.
Toco
Range: 0-100 relative units
Resolution: 1 count
Accuracy:
±1% ±1 relative unit
Leakage:
10 µ A @ 264 VAC applied
to transducer
Isolation: >4 kV RMS, Type BF
applied part
Transducers
Doppler Ultrasound FHR (M1356A)
Weight: 185 g (6.5 oz)
Cable length: 2.5m (8ft 2in)
Diameter: 75mm (2.95in)
Latex free
External Toco (M1355A)
Transducer Weight: 180 g (6.3 oz)
Cable length: 2.5m (8ft 2in)
Diameter: 75mm (2.95in)
Latex free
Remote Event Marker (15249A)
Length: 2.8m (9ft 2in)
1.USA only
Parameter
Input
Connector
Resulting
Isolation
US BF BF
Toco BF BF
Type: Lead acid, rechargeable.
Operation: minimum 2-hours fully
charged at 10
o
C, longer
at higher temperature.
Recharge during
monitoring:
Maximum 14 hours full
recharge during monitor-
ing (typical 11 hours).