Ramp-Up Transition from
Stimulation to Expression Phase.
When switching from Stimulation
to Expression Phase, the
Symphony 2.0 Program Card
increases the vacuum setting
by adding 3 bars to the last
Stimulation setting in a stair-
stepped fashion. The 3 additional bars and increased
vacuum setting occur gradually over a 4-second period.
The ramp-up transition eliminates potential discomfort of
an immediate vacuum setting increase.
44
Use Let-Down Button to Switch
Back to Stimulation Phase.
The Symphony pump can switch
from Stimulation to Expression
Phase in one of two ways: 1)
automatically after 2 minutes, or
2) manually through pressing the
let-down button. With the Symphony 2.0 Program Card,
the let-down button can also be used to switch back to
Stimulation Phase. For more information on this feature,
please review the Symphony 2-Phase Pumping Protocol
information in this booklet.
Note: Stimulation Phase is always limited to two minutes. If milk
ejection occurs before the 2 minute Stimulation Phase is over, press
the let-down button and continue pumping in Expression Phase
55
Program Remembers Last
Stimulation Vacuum Setting.
When switching back to
Stimulation Phase, the Symphony
2.0 Program Card remembers the
last Stimulation Phase vacuum
setting prior to switching to
Expression Phase and returns to
this setting in Stimulation Phase.
11
44 Vacuum Setting.
When switching from Stimulation to
Expression, the pump automatically
increases the vacuum setting by
adding 3 Bars to the last
Stimulation setting. *This vacuum
level can be increased by using the
Vacuum Regulator Knob.
Symphony
®
2.0 Program Card
The Symphony 2.0 Program Card is the second genera-
tion program card. It is programmed to operate the
2-Phase Expression similar to the Original Symphony
Program Card. Additionally, there are some new features
available with the Symphony 2.0 Program Card:
11
Program Card Name Displayed
on LCD Display.
The Symphony 2.0 Program Card
displays the name of the card
(Symphony 2.0) on the LCD display
when the pump is powered off and
connected to A/C power supply.
When batteries are used to power
the Symphony, the LCD Display will
remain blank until powered on.
22
Minimum Start Vacuum Setting.
The Symphony 2.0 Program
Card will start the pump in the
Stimulation Phase. The initial
vacuum setting when powering on
the pump is visually indicated with
3 drops on the LCD screen. The
vacuum setting can be increased
or decreased by using the Vacuum
Regulator Knob.