A&D GP-61K Building Set User Manual


 
60
The following commands can not be used during data storage.
Q Query command for weighing data.
S Query command for stable weighing data.
SI Query command for weighing data.
SIR Query command for continuous weighing data.
Setting the function table
Parameter settings for each output mode are as follows:
Item
Mode
Data output
mode
Auto print
polarity,
difference
Data memory
function
Interval
time
Key mode prt 0 Not used data 2
Auto print mode A prt 1 data 2
Auto print mode B prt 2
ap-a 0-2
ap-b 0-2
data 2
Not used
Interval memory mode prt 3 Not used data 2 int 0-8
Set each item, depending on the situation, as follows:
No d-no 0 No 5-td 0
Data number
Yes d-no 1 Time only 5-td 1
No 5-id 0 Date only 5-td 2 ID number
Yes 5-id 1
Time and date
Both 5-td 3
With 5-td 1, 5-td 2 or 5-td 3 selected, the amount of data to be stored is 100 sets.
Enabling the data memory function
1 Press and hold the SAMPLE key until ba5fnc is displayed, then release the key.
2 Press the SAMPLE key several times to display dout .
3 Press the PRINT key.
4 Press the SAMPLE key three times to display data 0 .
5 Press the RE-ZERO key to display data 2 .
6 Press the PRINT key to store the setting.
7 Press the CAL key to return to the weighing mode.