A&D GP-30KS Building Set User Manual


 
43
Interval memory mode
The weighing data is periodically stored in memory.
Required setting dout prt 3 Interval memory mode
dout data 2 Data memory function is used.
Stores weighing data.
dout int Interval time
Optional setting dout 5-td1, 2, or 3 Adds the time and date.
Example “For periodical weighing without a computer command and to output
all of the data, to a computer, at one time”
9-6 Description of the Item “Data format”
A&D standard format 5if type 0
This format is used when the peripheral equipment can receive the A&D format. If an AD-8121B is
used, set the printer to MODE 1 or 2.
This format consists of fifteen characters excluding the terminator.
A header of two characters indicates the balance condition.
The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus
sign is used.
The unit, consisting of three characters, follows the data.
DP (Dump print) format 5if type 1
This format is used when the peripheral equipment can not receive the A&D format. If an AD-8121B
is used, set the printer to MODE 3.
This format consists of sixteen characters excluding the terminator.
A header of two characters indicates the balance condition. No overload header is used.
The polarity sign is placed before the data, with spaces in place of leading zeros, if the data is
not zero or overloaded.
The unit, consisting of three characters, follows the data.