A&D GP-100KS Building Set User Manual


 
88
GP-60K
GP-60KS
GP-61K
GP-61KS
GP-100K
GP-100KS
GP-102K
Weighing capacity 61 kg 61 kg 101 kg 101 kg
Maximum display (See bottom for )
61.084 kg 61.0084 kg 101.084 kg 101.08kg
61.009 kg
Minimum weighing value (1 digit) 1 g 0.1 g 1 g 10 g/1 g
Repeatability (Standard deviation) 0.7 g 0.2 g 1 g 5 g/1 g
Linearity
±1 g ±0.5 g ±2 g ±10 g/±2 g
Stabilization time (typical at FAST)
Approx.1.5 seconds
Sensitivity drift
(10°C-30°C/50
°
F-86
°
F )
∗∗
±6ppm/°C ±10 ppm/°C
±5 g ±3 g ±10 g
Accuracy right after calibration using
the internal mass. ∗∗∗
The value above is to the weighing capacity.
Operating environment 5°C to 40°C (41°F to 104°F)
85%RH or less (No condensation)
Display refresh rate 5 times/second or 10 times/second
Minimum unit mass 1 g 0.1 g 1 g 10 g Counting
mode
Number of samples 10, 25, 50 or 100 pieces
Minimum 100% reference mass
100 g 10 g 100 g 1000 g Percent
mode
Minimum 100% display 0.01%, 0.1%, 1% (Depends on the reference mass stored.)
Interface (Provided as standard) RS-232C with Windows Communication Tools WinCT
External calibration weight 30 kg
40 kg
50 kg
60 kg
60 kg
80 kg
100 kg
Weighing pan 384 x 344 mm 386 x 346mm
Dust and water protection Complying with IP65
Sensor used Super Hybrid Sensor (SHS)
AC adapter Confirm that the adapter type is correct for the local voltage
and power receptacle type
Power consumption Approx. 14VA (supplied to the AC adapter )
External dimensions (WxDxH, mm) 372 x 615 x 130
(GP-60KS/ GP-61KS:
344 x 442 x 130 ) ∗∗∗∗
373 x 615 x 130
(GP-100KS: 346 x 443 x 130 )
∗∗∗∗
Net weight Approx. 17 kg Approx. 18 kg
The GP-102K allow weighing using a precision range, even with a heavy tare placed on the
pan. (Smart range function)
∗∗ When automatic self calibration is not used, at 10°C to 30°C (50°F to 86°F)
∗∗∗ Accuracy right after calibration using the internal mass in good ambient conditions (within the
temperature range of 10°C to 30°C (50°F to 86°F) with no abrupt changes in temperature or
humidity, no drafts, no effect by magnetic fields or static electricity).
Check the internal mass periodically as described in “17. MAINTENANCE”.
∗∗∗∗ The size for the GP-60KS/61KS/100KS is that of the main unit. The display size is 260x164x202.
The cable length, for the separate display type balances (GP-60KS/61KS /100KS), is 3 meters.