3M 1292 Model Vehicle User Manual


 
3M™ Attest™ 1292 Rapid Readout Biological Indicator
ProductDescription
The 3M™Attest™ 1292Rapid Readout Biological Indicator (brown cap) is a dual readout biological indicator system specifically designed for rapid and reliable monitoring of steam
sterilization processes when used in conjunction with the 3M™Attest™ 290Auto‑reader or the 3M™Attest™ Auto‑reader390.
The Attest™ 1292Rapid Readout Biological Indicator detects the presence of Geobacillus stearothermophilus by detecting the activity of alpha‑glucosidase, an enzyme present within
the organism. The presence of the enzyme is detected by reading fluorescence produced by the enzymatic breakdown of a non‑fluorescent substrate. This creates a fluorescence
change, which is detected by the auto‑reader. A fluorescence change indicates a steam sterilization processfailure.
The Attest™ 1292Rapid Readout Biological Indicator also detects the presence of G. stearothermophilus organisms by a visual color change reaction. Biochemical activity of the G.
stearothermophilus organism produces acid by‑products that cause the media to change color from purple to yellow. A visual pH color change also indicates a steam sterilization process
failure. Due to the high sensitivity of the 3‑hour fluorescent results, however, there is no advantage to incubating the Attest™ 1292Rapid Readout Biological Indicator beyond 3hours.
Indications
Use the Attest™ 1292Rapid Readout Biological Indicator tomonitor:
1. 250°F (121°C) gravity steam sterilizationcycles.
2. 270°F (132°C) vacuum assisted steam sterilizationcycles.
Contraindications
None.
Warning
There is a glass ampoule inside the plastic vial of the biological indicator. To avoid the risk of serious injury from flying debris due to a ruptured biologicalindicator:
•
Allow the biological indicator to cool for the recommended time period before crushing. Crushing or excessive handling of the biological indicator before cooling may cause the
glass ampoule toburst.
•
Wear safety glasses and gloves when removing the biological indicator from thesterilizer.
•
Wear safety glasses when crushing the biologicalindicator.
•
Handle the biological indicator by the cap when crushing andtapping.
•
Do not use your fingers to crush the glassampoule.
•
Do not roll the biological indicator between fingers to wet the sporestrip.
Precautions
Do not use the Attest™ 1292Rapid Readout Biological Indicator tomonitor:
1. 250°F (121°C) vacuum assisted steam sterilizationcycles.
2. 270°F (132°C) gravity steam sterilizationcycles.
3. Dry heat, chemical vapor, ethylene oxide or other low temperature sterilizationprocesses.
MonitoringFrequency
Follow facility Policies and Procedures which should specify a biological indicator monitoring frequency compliant with professional association recommended practices and/or
national guidelines and standards. As a best practice and to provide optimal patient safety, 3M recommends that every steam sterilization load be monitored with an appropriate
biologicalindicator.
Directions forUse
1. Identify the Attest™ 1292biological indicator by writing the sterilizer and load ID number and processing date on the indicator label. Do not place another label or indicator tape on
the vial or on thecap.
2. Place the Attest™ 1292biological indicator in an appropriate process challenge device according to recommended practices. Do not place the Attest™ 1292biological indicator in
direct contact with a chemical indicator. Fluorescent residue could transfer to the biological indicator and affect theresult.
Appropriate process challenge devices for loadscontaining:
a. Fabric packs run at 250°F (121°C) for ≥40minutes in a gravity displacement cycle or at 270°F (132°) for ≥4minutes in a vacuum assistedcycle:
Place an Attest™ 1292biological indicator in an AAMI 16towelpack.
b. Wrapped hard goods run at 250°F (121°C) for ≥20minutes in a gravity displacementcycle:
Place an Attest™ 1292biological indicator in a wrapped hard good item (e.g., instrument set) from theload.
c. Wrapped hard goods run at 270°F (132°C) for ≥4minutes in a vacuum assistedsterilizer:
Place an Attest™ 1292biological indicator in a wrapped hard good item (e.g., instrument set) from theload.
Place an Attest™ 1292biological indicator in an AAMI 16towelpack.
d. Single wrapped hard goods run at 270°F (132°C) for ≥4minutes in an express vacuum assistedcycle:
Place an Attest™ 1292biological indicator in a single wrapped hard goods item (e.g., instrument tray) from the load. Check recommendations of the manufacturer before
processing items with lumens or porousmaterials.
e. Unwrapped hard goods with no porous items run at 250°F (121°C) in a gravity displacement cycle for ≥ 15minutes or 270°F (132°C) for ≥ 3minutes in a vacuum
assistedsterilizer:
Place an Attest™ 1292biological indicator in unwrapped instrument hard good item (e.g., instrument set) from the load. AAMI suggests placing a biological indicator in an
emptytray.
f. Unwrapped metal hard goods with porous items run at 270°F (132°C) for ≥ 4minutes in a vacuum assistedsterilizer:
Place an Attest™ 1292biological indicator in an unwrapped instrument hard goods item (e.g., instrument tray) from the load, including porous items. AAMI suggests placing a
biological indicator in an empty tray but include porous items ifapplicable.
g. Container systems run at 250°F (121°C) for ≥ 40minutes in a gravity cycle or 270°F (132°C) for ≥ 4minutes in a vacuum assistedcycle:
Place the Attest™ 1292biological indicator in the areas determined by product testing to provide the greatest challenge to the sterilizationprocess.
3. Place the test tray or package in the most challenging area for the sterilant. This is typically on the bottom shelf, near the door and over thedrain.
4. Process the load according to recommendedpractices.
5. After completion of the cycle, fully open the sterilizer door for a minimum of 5minutes prior to removing the Attest™ 1292biologicalindicator.
6. When the biological indicator is not contained in a test pack or other heat absorbing packaging material, remove the biological indicator from the sterilizer and allow it to cool for an
additional 10minutes prior tocrushing.